Pre-installation requirements of water chiller

Notice:

BLK: Serial number of water chiller

X:Cooling capacity(The cooling capacity when ambient temperature is 25,water chiller supply

water temperature for 25.)

FF: Divided air-cooling type

YF: Integrated air-cooling type

YS: Integrated water-cooling type

X: Temperature stability

Pre-installation requirements of BLK-xFF

1. Location

1.1. The heat from the equipment and the water chiller itself is removed through the cooling

fins of the condenser and fan. The maximum heat can reach __kw.

1.2. The unit should be located in a laboratory or clean industrial environment. Never place

the unit in a location where excessive heat, moisture, or corrosive materials are present.

1.3. The unit has an air-cooled refrigeration system. It must be positioned so the air intake

and discharge are not impeded.

1.4. The installation of outdoor and indoor locations of the best distance is not more than 4

meters, reducing the length of pipeline will help reduce cooling loss. the favorable distance

between the indoor and outdoor set is no more than 4m

1.5. Outdoor location of the installation should be no higher please see to it that position of

the outdoor set should not be higher ?than the indoor machine; otherwise it will affect the

life span of the compressor.

1.6. Outdoor installation of the wall need to load-bearing walls (such as cement structure),

installation location concern should help the installation, maintenance and repair.

1.7. In the indoor unit installed near the wall at the top of 200 mm from the ground, the need

to fight more than a Φ_mm hole to the outdoors.

1.8. Outside ambient temperature:≤40℃.

Type
BLK-2FF-B

BLK-5FF-B/R

BLK-8FF-B/R

BLK-12FF-B/R

BLK-18FF-B/R

BLK-25FF-B/R

Max heat

3KW

7KW

10/11KW

15KW

23KW

31KW

Φ_mm

50mm

60/80mm

60/100mm

80/100mm

120mm

150/200mm

2. Power supply

2.1Power supply lead section≥2.5mm2, earth lead section ≥1.5mm2.

2.2For power supply, fluctuation of voltage ≤10%, fluctuation of frequency ≤0.5%, grounding

resistance of ground wire ≤10Ω.

Attention〉It must install the protective grounding wire.

2.3The length of unit power cable is 3 meters.

 

Pre-installation requirements of BLK-xYF

1. Location

1.1. The heat from the equipment and the water chiller itself is removed through the cooling

fins of the condenser and fan. The maximum heat can reach xkw.

Max heat BLK-2YF BLK-5YF BLK-8YF BLK-12YF BLK-18YF BLK-25YF BLK-48YF
x 3KW 7KW 10KW 15KW 22KW 31KW 60KW

 

 

1.2. The unit should be located in a laboratory or clean industrial environment. Never place

the unit in a location where excessive heat, moisture, or corrosive materials are present.

1.3. The unit has an air-cooled refrigeration system. It must be positioned so the air intake

and discharge are not impeded. The requirement of the installation position of the water

chiller is shown as figure.

1.4. Ambient temperature: 5~35℃.

2. Power supply

2.1. Power supply lead section≥2.5mm2, earth lead section ≥1.5mm2.

2.2. For power supply, fluctuation of voltage ≤10%, fluctuation of frequency ≤0.5%, grounding

resistance of ground wire ≤10Ω.

Attention 〉It must install the protective grounding wire.

2.3. The length of unit power cable is 3 meters.

 

Pre-installation requirements of BLK-xYS

1. Location

1.1. The heat from the equipment and the water chiller itself is removed through the PWT

evaporator and cooling water. The maximum heat can reach xkw.

Max heat

BLK-5YS

BLK-8YS

BLK-12YS

BLK-25YS

x

7KW

11KW

15KW

31KW

1.2. The unit should be located in a laboratory or clean industrial environment. Never place the

unit in a location where excessive heat, moisture, or corrosive materials are present.

1.3. The unit has a water-cooled refrigeration system. It must connect external cooling water,

and the sewer.

Cooling water

Type

 

10℃

15℃

20℃

25℃

30℃

BLK-5YS

Water flow

4.2L/min

8.5 L/min

12.2 L/min

19.2 L/min

28.5 L/min

BLK-8YS

Water flow

6.3 L/min

13.1 L/min

19.6 L/min

28.1 L/min

42.3 L/min

BLK-12YS

Water flow

10.5 L/min

21.8 L/min

32.6L/min

46.8 L/min

70.5 L/min

BLK-25YS

Water flow

21 L/min

43.6L/min

65.2L/min

93.6 L/min

140 L/min

1.4. Ambient temperature: 5~35℃.

2. Power supply

2.1. Power supply lead section≥4.0mm2, earth lead section ≥2.5mm2.

2.2. For power supply, fluctuation of voltage ≤10%, fluctuation of frequency ≤0.5%, grounding

resistance of ground wire ≤10Ω.

AttentionIt must install the protective grounding wire.

2.3. The length of unit power cable is 3 meters.

 

 

 

 

 

 